Conference Publication Details
Mandatory Fields
O'Keeffe, S,Grattan, M,Hounsell, A,McCarthy, D,Woulfe, P,Cronin, J,Lewis, E,Jaroszewicz, LR
FIFTH EUROPEAN WORKSHOP ON OPTICAL FIBRE SENSORS
Radiotherapy dosimetry based on plastic optical fibre sensors
2013
January
Published
1
()
Optional Fields
Optical fibres sensors radiotherapy brachytherapy dosimetry PMMA optical fibres
The use of a PMMA based plastic optical fibre in radiotherapy dosimetry is presented. The optical fibre tip is coated with a scintillation material, terbium-doped gadolinium oxysulfide (Gd2O2S:Tb), which fluoresces under ionising radiation. The emitted signal penetrates the fibre and propagates along the fibre where it is remotely monitored using a fluorescence spectrometer. The results demonstrate good repeatability, with a maximum percentage error of 0.59% and the response is independent of dose rate.
10.1117/12.2025864